Limit values |
|
Maximum capacity |
220 g |
Maximum capacity, fine range |
82 g |
Readability |
0.1 mg |
Readability, fine range |
0.01 mg |
Repeatability (at nominal load) 1) |
0.08 mg (200 g) |
Repeatability (at low load) 1) |
|
Linearity deviation |
0.2 mg |
Sensitivity offset 2) |
0.8 mg (200 g) |
|
|
|
|
Typical values |
|
Repeatability (at nominal load) 1) |
0.07 mg (200 g) |
Repeatability (at low load) 1) |
0.015 mg (20g) |
Minimum weight (according to USP) |
45 mg |
Minimum weight (@U=1%, 2 sd) |
3 mg |
Settling time |
4s / 8s |
Sensitivity offset 2) |
0.6 mg (200g) |
|
|
|
|
Dimensions |
|
Weighing pan dimensions (mm) |
ø 80 |
Balance dimensions LxWxH (mm) |
358x247x331 |
1) Standard deviation (sd)
2) Stability of sensitivity with FACT self-adjustment switched on